Vertical Laminar Flow Cabinets PLUG&PLAY models FL-1 and FL-2, 670-FL and 870-FL. The cabinets are removable. In both, the air flow passes through an H-14 HEPA filter (manufactured according to the EN-1822 standard) at a constant speed, traveling through the cabin in a laminar regime and in a vertical direction, generating a Class 100 sterile area.
(according to the American Federal Standard 209E and equivalent to the ISO Class 5 standard) and protecting the product from external contamination and cross contamination between samples.
Single-stage filtration system with a minimum efficiency of 99.995% for 0.3 µm particles at the top and with 100% extraction of the air flow to the outside.
Models |
670-FL |
870-FL |
FL-1 |
FL-2 |
Internal Sizes (Width/Depth/High) |
575x560x630 |
775x560x740 |
710x556x610 |
1110x556x610 |
External Sizes (Width/Depth/High) |
600x600x930 |
800x600x1178 |
800x600x1137 |
1200x600x1137 |
Filter columns |
– |
– |
– |
– |
Filter number |
1 |
1 |
1 |
1 |
Fans |
1 |
1 |
1 |
1 |
Average volume of treated air/hour |
– |
– |
– |
– |
Average face velocity |
0.45 m/s |
0.45 m/s |
0.45 m/s |
0.45 m/s |
Internal volume of the cabinet |
0.181m3 |
0.287m3 |
0.236m3 |
0.376m3 |
Renewals inside the cabinet/min. |
– |
– |
– |
– |
Total electrical power consumption |
102 W |
226 W |
226 W |
226 W |
Total electrical power consumption |
110-220 V / 50-60 Hz |
110-220 V / 50-60 Hz |
110-220 V / 50-60 Hz |
110-220 V / 50-60 Hz |
LED light intensity |
900Lux / LED 16 W |
900Lux / LED 16 W |
900Lux / LED 16 W |
900Lux / LED 16 W |
UV Lamp |
1x15W |
1x15W |
1x15W |
1x15W |
Sound level |
48 dB |
50 dB |
50 dB |
50 dB |
Weight |
65 Kg |
92 Kg |
112 Kg |
142 Kg |
|
0.28 m3 |
0.42 m3 |
0.74 m3 |
0.95 m3 |